Configuration
Menu item
Action
Choices
Output current
during error message
Select ON (22 mA for
error message) or OFF
using
keys.
Press
enter
to confirm.
ON/
OFF
Output current during
Sensoface messages
OT1: FACE 22 mA
Select ON or OFF using
keys.
Confirm by pressing
enter
ON/
OFF
Output current
during HOLD
LAST: During HOLD the
last measured value is
maintained at the output.
FIX: During HOLD a value
(to be entered) is main‑
tained at the output.
Select using
Press
enter
to confirm.
LAST
/FIX
Output current for
HOLD FIX
Only with FIX selected:
Enter current which is to
flow at the output during
HOLD
Enter value using
keys.
Press
enter
to confirm.
04.00...22.00 mA
(
21.00 mA
)
HOLD active
21
4
HOLD active
Output current
[mA]
Output signal for HOLD
FIX setting = 21.0 mA
Output signal for HOLD
LAST setting
Output signal during HOLD:
5