Error
Info text
(is displayed in case of
fault when the Info key is
pressed)
Problem
Possible causes
ERR 10
CONDUCTANCE TOO HIGH
Measuring range of
conductance exceeded
> 3500 mS
ERR 11
CONDUCTIVITY RANGE
CONCENTRATION RANGE
SALINITY RANGE
Display range limits
exceeded
Cond > 1999 mS/cm
> 99.99 S/m
< 1 ohm * cm
Conc > 99,99 %
SAL > 45.0 ‰
ERR 13
TEMPERATURE RANGE
Temperature range limits
exceeded
Connect the sensor, check the
sensor cable and replace if neces-
sary, check the sensor connection,
adjust the parameter settings.
ERR 15
SENSOCHECK
Sensocheck
ERR 60
OUTPUT LOAD
Load error
Check the current loop, deactivate
unused current outputs.
ERR 61
OUTPUT 1 TOO LOW
Output current 1
< 3.8 mA
ERR 62
OUTPUT 1 TOO HIGH
Output current 1
> 20.5 mA
ERR 63
OUTPUT 2 TOO LOW
Output current 2
< 3.8 mA
ERR 64
OUTPUT 2 TOO HIGH
Output current 2
> 20.5 mA