Specifications
Output 2
For version with 2nd
current output only
Current loop, 4 ... 20 mA, floating, reverse polarity protected
Supply voltage
14 ... 30 V
Process variable *
Conductivity, resistivity, concentration, salinity or temperature
Characteristic *
Linear, bilinear, or logarithmic
Overrange *
22 mA in the case of error messages
Output filter *
PT
1
filter, time constant 0 ... 120 s
Measurement error
1)
< 0.25 % of current value + 0.05 mA
Start/end of scale *
Configurable within selected range
Bilinear: Vertex X/Y *
Configurable within selected range
Real-time clock
Different time and date formats selectable
Power reserve
> 5 days
Display
LC display, 7-segment with icons
Main display
Character height approx. 22 mm, unit symbols approx. 14 mm
Secondary display
Character height approx. 10 mm
Text line
14 characters, 14 segments
Sensoface
3 status indicators (friendly, neutral, sad face)
Mode indicators
meas, cal, conf, diag
Further icons for configuration and messages
Alarm indication
Display blinks, red backlighting
Keypad
Keys: meas, menu, info, 4 cursor keys, enter
HART communication (TAN)
HART version 6
Digital communication by FSK modulation of output current 1
Device identification, measured values, status and messages,
parameter setting, calibration, records
FDA 21 CFR Part 11
Access control by editable passcodes
Logbook entry and flag via HART in the case of configuration changes
Message and logbook entry when enclosure is opened