Specifications
COND input
Input for 2-/4-electrode sensors and Memosens sensors
Measuring range
2-el. sensors
0.2 µS · c … 200 mS · c
4-el. sensors
0.2 µS · c … 1000 mS · c
(Conductance limited to 3500 mS)
Display ranges
Conductivity
0.000 ... 9.999 µS/cm
00.00 ... 99.99 µS/cm
000.0 ... 999.9 µS/cm
0000 ... 9999 µS/cm
0.000 ... 9.999 mS/cm
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
0.000 ... 9.999 S/cm
00.00 ... 99.99 S/cm
Resistivity
00.00 … 99.99 MΩ · cm
Concentration
0.00 ... 9.99 %
Salinity
0.0 ... 45.0 ‰ (0 ... 35 °C / 32 ... 95 °F)
Response time (T90) Approx. 1 s
Measurement error
1,2,3)
< 1 % meas. val. + 0.4 µS · c
Temp compensation
*
(OFF)
Without
(LIN)
Linear characteristic 00.00 ... 19.99%/K
(reference temp user-defined)
(NLF)
Natural waters to EN 27888
(reference temp 25 °C / 77 °F)
(NACL)
Ultrapure water with NaCl traces (0 ... 120 °C / 32 ... 248 °F),
reference temp 25 °C / 77 °F
(HCL)
Ultrapure water with HCl traces (0 ... 120 °C / 32 ... 248 °F),
reference temp 25 °C / 77 °F
(NH3)
Ultrapure water with NH3 traces (0 ... 120 °C / 32 ... 248 °F),
reference temp 25 °C / 77 °F
(NaOH)
Ultrapure water with NaOH traces (0 ... 120 °C / 32 ... 248 °F),
reference temp 25 °C / 77 °F
Concentration
determination
-01- NaCl
0 – 26 wt% (0 °C / 32 °F)
... 0 – 28 wt% (100 °C / 212 °F)
-02- HCl
0 – 18 wt% (-20 °C / -4 °F)
... 0 – 18 wt% (50 °C / 122 °F)
-03- NaOH
0 – 13 wt% (0 °C / 32 °F)
... 0 – 24 wt% (100 °C / 212 °F)
-04- H
2
SO
4
0 – 26 wt% (-17 °C / 1.4 °F)
... 0 – 37 wt% (110 °C / 230 °F)
-05- HNO
3
0 – 30 wt% (-20 °C / -4 °F)
... 0 – 30 wt% (50 °C / 122 °F)