Configuration
Menu item
Action
Choices
CIP / SIP
Cleaning cycles
Select ON or OFF using
keys.
Activates/deactivates log‑
ging in extended logbook
(TAN SW‑A003).
Press
enter
to confirm.
ON/
OFF
Sterilization cycles
Select ON or OFF using
keys.
Activates/deactivates log‑
ging in extended logbook
(TAN SW‑A003).
Press
enter
to confirm.
ON/
OFF
Logging the cleaning and sterilization cycles with connected sensor
helps measuring the load on the sensor.
Suitable for biochemical applications (process temp approx.
0 ... 50 °C / 32 ... 122 °F, CIP temp > 55 °C / 131 °F,
SIP temp > 115 °C / 239 °F).
Note:
A CIP or SIP cycle is only entered into the extended logbook (TAN
SW‑A003) 2 hours after the start to ensure that the cycle is complete.
With Memosens, an entry is also made in the sensor.