maxium® smart C Electrosurgery Unit with maxium® smart Beam
60
Revision 1
10.5.5
Deleting programs
To delete a program, proceed as follows:
•
Press the symbol
to open the
Setup
•
Press
Program settings
•
Press
Delete Program
•
Find the program to be deleted and mark it
•
Confirm by tapping the
check mark
or
Delete
program
•
Confirm again by tapping the
check mark
The base program and programs with the “padlock”
icon can be neither deleted nor overwritten. Changes
can be saved only under a new program name, see
section 10.5.4 “Saving a program under a new name”,
page 59.
10.5.6
SWAP® mode (optional)
The SWAP® mode allows switching to and fro between two pre-set programs by using the foot switch or
the finger switch on electrode handle.
This new function allows the operator to use different settings during the operation, without the need for
switching to be made by a non-sterile person directly at the maxium® smart C. It is useful where there is a
need to change any settings of the unit in the course of a surgery.
Switching from a user program to the associated parallel program is possible by using the special double-
pedal foot switches with SWAP® mode or by simultaneously activating both finger switches of any two-key
electrode handle. This function cannot be used with electrode handles in which the two keys are combined
into a rocker switch.
The maxium® smart C confirms the switching by a multiple acoustic signal.
For using the SWAP® mode, the following conditions must be created:
First, for the active program a SWAP® program must
be defined.
To this purpose, an already existing program can be
used, otherwise first a program must be generated
and saved, for which purpose usually the program
will be used which it is to be assigned to afterwards.