maxium® smart C Electrosurgery Unit with maxium® smart Beam
126
Revision 1
17.3
Technical Data for maxium® smart Beam
Description
maxium® smart Beam
Power supply
via the maxium® smart C
Argon supply
Only approved for MABS pressure reducer.
Fuse
M1.6A, internal
Power consumption
max. 40 W
Protection class
I
Classification acc. to Medical
Devices Act
II b
Argon gas flow for CUT
0.1–12 l/min ± 20%
Argon gas flow for COAG
0.1–12 l/min ± 20%
Leakage currents LF and HF
according to IEC 60601, Part 1
(to be tested together with the HF generator of the maxium® smart C)
Type (Application Part)
CF; defibrillation-proof
Operating mode
INT 10s/30s
Weight
3.5 kg
Radio interference suppression
See section 17.2. The maxium® smart C meets the requirements of
IEC 6060112 also in conjunction with the maxium® smart beam, but with the
restrictions pursuant to IEC 6060122 Section 202.6.1.
Dimensions
Width
Height
Depth
390 mm
108 mm
425 mm
Environmental conditions for
transport and storage
Ambient temperature
Relative humidity
Air pressure
-25°C to +70°C (-13°F to +158°F)
10–100%
500–1,060 hPa
Environmental conditions for
operation
Ambient temperature
Relative humidity
Air pressure
+10°C to +40°C (50°F to +104°F)
30–75%
700–1,060 hPa
Compliant with 93/42/EEC