maxium® smart C Electrosurgery Unit with maxium® smart Beam
110
Revision 1
13.3
Input Feedback
The maxium® smart C will not accept illegal input or settings. A message will appear. In order to resolve the
conflict, proceed as described in the table.
Message
Description
This current is not available due to selection
of a Baby NE
An attempt was made to select a monopolar current excluded by
the setting Baby NE in the
Setup
menu, see section 10.6.4, page 65.
This current is not available due to selection
of a Mega Soft NE
An attempt was made to select a monopolar current excluded by
the setting Mega Soft-NE in the
Setup
menu, see section 10.6.4,
page 65.
Single-piece NE is prohibited by Setup,
hence NE error!
The unit has detected connection of a single-piece neutral
electrode, although this in not permissible in accordance with the
change of the setting in the
Setup
menu. Activation of monopolar
power is not possible in this fashion. If the unit is to be used (again)
with single-piece electrodes, the setting must be reverted in the
Setup
menu, see section 10.6.4, page 65.
The incompatible current on the same
output port was disabled
The alternating activation of bipolar cutting current and bipolar
coagulation current with automatic function on the same output is
not possible. Either current can be selected. Selection of one
current will, however, automatically lead to deselection of the
other.