maxium® smart C Electrosurgery Unit with maxium® smart Beam
36
Revision 1
6.3.5
Mains connection
The maxium® smart C is equipped with a wide-range power supply, so that the nominal
value of the mains voltage can be from 100 V to 240 V without need for switching the
network connection. The power connector (24, Fig. 6-1) also comprises the fuse
protection of the unit with two fuses T 6.3 AH, 250 V.
6.4
Controls
For control of the maxium® smart C, a touchscreen is provided next to the
ON/OFF
6.4.1
ON/OFF key
After proper preparation and installation of the maxium® smart C (see national
installation and operation regulations and national rules for fire and explosion
protection), in the designated areas the unit can be set into operational condition by
pressing the
ON/OFF
key (6, Fig. 6-1).
The maxium® smart C is set to display the last used program after each switching on. However, in the
Service
menu, the base program can be defined as autostart program (factory default, see section 10.6.7
If the automatic function of the bipolar coagulation
Auto Start
is to be used, for safety reasons it must be
activated after each switching on of the maxium® smart C.
If any setting of the program which was active during the last switching off has been modified, there
appears a notification that the program will be started with the default settings. This is done for safety
reasons in order to alert the user to the fact that the settings of the program on the screen are not the
same as during the last switching off.
The ON/OFF key will not disconnect the unit from the grid!
To disconnect the unit from the grid, the plug of the connecting cable must be pulled – at the unit or from
the outlet.
6.4.2
Touchscreen
For control of the maxium® smart C, the corresponding menu items are controlled directly via the
touchscreen.