maxium® smart C Electrosurgery Unit with maxium® smart Beam
66
Revision 1
If a SWAP program is to be created using a program with a specific NE, for safety reasons the specific NE is
applied for the other program as well. This NE is then displayed in the other program as “modified”.
Channels with current types that may potentially not be allowed are deactivated.
When using third-party accessories, always observe the manufacturer's instructions, and especially safety
notes and information concerning the parameters to be used.
10.6.5
Argon mode
This menu can be used to define whether the argon currents and the functions of the maxium® smart Beam
are to be displayed in the menu levels even if no maxium® smart Beam is connected. If the argon mode is
not activated, in some of the menus some functions will be eliminated, making the menus easier to read.
Thus the currents with argon beamer are not displayed in the list of selectable currents if beamer operation
is deselected. If a maxium® smart Beam is connected to the unit, then argon mode will be activated
automatically. By default, the unit is set not to display the beamer-relevant functions if no beamer is
connected. This is a global setting which, however, does not apply to the base program, where the beamer
parameters will always be displayed. There, deactivation of beamer operation causes only a temporary
hiding until the unit is turned off and on again.
10.6.6
Hotline
Under
Hotline
, contact information for direct help in case of questions about or problems with the
maxium® smart C can be found.
10.6.7
Service
The
Service
menu provides access to the functions which are not needed in the clinical operation of the
maxium® smart C. They are therefore accessible for technicians and service only. This includes, inter alia,
the following functions:
•
Adjustment of the gas cylinder filling level
•
Beamer parameters: Setting of the alarm threshold and other argon-related parameters.
•
Adjustment of the starting process: Autostart with base program or previously used program
•
Special use of the base program
•
Program settings: Locking and unlocking of programs
•
Unit status: Display of the hardware and software configuration, detection of peripherals.
•
Reset to factory default settings
Since custom settings can be deleted irreversibly (e.g. Reset to factory defaults will delete all personal user
settings) in this menu, it is password-protected. Details are described in the Service Manual for the
maxium® smart C by Gebrüder Martin.
10.6.8
Interface Language
Selection of the user interface language: The texts of the menus, functions, and fields, as well as the
messages in the status line, can be displayed in various national languages.