maxium® smart C Electrosurgery Unit with maxium® smart Beam
54
Revision 1
10.3
Setting the Values for the Working Channels
The channel fields are used for
•
selecting the HF output power, page 54,
•
selecting the current, page 54,
•
in the case of argon-supported currents also for setting the gas flow, page 54.
•
setting of the effect, page 55,
•
selection of the activation type/foot switch, page 55.
10.3.1
Setting the HF Output Power
Below the number for the power setting there is a
horizontal bar indicating the power relative to its
maximum value. This indicator is not linear.
After pressing of the power value to be adjusted,
a red frame appears around the power value and a
window is opened.
The power can now be adjusted using the slider
and/or the
+
/
-
buttons. Any adjustment can be
confirmed by tapping the check mark or canceled
by tapping the cross.
10.3.2
Selecting the current
The maxium® smart C is equipped with various current types for monopolar and bipolar applications,
for cutting and coagulation, depending on the application and instrument.
The individual icons indicate the differences in performance in form of the dots (more dots stand for
increased coagulation) and the type of instrument to be used (e.g. cutting loop of a resectoscope).
After pressing the current type to be adjusted, a red
frame appears around the current type, and a window
is opened. Here, another current type can be selected
by pressing the icon (a black frame appears).
The selection can be confirmed by tapping the check
mark or canceled by tapping the cross.
Each current type has a default value for the power
and effect stored, which will be used when the
selection of the current type is confirmed. If these
default values are changed, they must be sub-
sequently changed separately.