![KLS Martin maXium smart C Instructions For Use Manual Download Page 106](http://html1.mh-extra.com/html/kls-martin/maxium-smart-c/maxium-smart-c_instructions-for-use-manual_1979004106.webp)
maxium® smart C Electrosurgery Unit with maxium® smart Beam
106
Revision 1
12
Cleaning and Disinfection
The responsibility for cleaning and disinfection lies with the operator/user of the product. Be sure to
observe your local cleaning and hygiene regulations!
The manufacturer recommends wiping disinfection according to the instructions of the manufacturer of the
disinfection agent.
Danger of serious injury when using defective accessories!
Using a unit that is not in perfect working order or using defective accessories can pose a hazard to
patients or users and compromise the proper functioning of the device.
•
Be sure to maintain the unit and its accessories in perfect operating condition.
•
Perform a visual inspection and check the system for insulation damage, cleanliness and integrity.
•
A unit or accessories found to be unfit for use must be withdrawn from service.
•
Never sterilize the unit!
•
Always disconnect the unit from the supply before cleaning and disinfecting it.
•
No liquid must be allowed to enter the unit. Therefore clean and disinfect it with care. Do
not
use
sprays
!
•
Never clean the unit with abrasive cleaners, disinfectants or solvents that could scratch the housing or
damage the device in other ways.
•
For surface cleaning and surface disinfection, use the procedure recommended by your hospital or use
another nationally recognized and approved method.
•
When disinfecting accessories (by surface disinfection or immersion), be sure to observe the
instructions of the manufacturer of the disinfectant regarding material compatibility, dosage and
exposure time.
•
Accessories not suitable for sterilization, such as the foot switch, must be cleaned with a disinfecting
detergent.
•
All disinfectant residues must be completely removed before putting the unit in operation.
•
If accessories must be disinfected or sterilized, be sure to observe the respective Instructions for Use.