In the socket module of the e-version, the sockets M are suitable
for instruments with 5-mm plug (coaxial circular connector with
5-mm inner conductor).
The following cables can be connected here:
•
Connector cable with coaxial circular connector with 5-mm
inner conductor
•
Connecting cable with US 3-pin plug
•
Instrument with a plain 4-mm connector (banana plug)
For this plug, use the right socket. Insertion into either of the
neighboring control contacts will cause malfunction!
6.2.2
Bipolar connections
The socket of the socket module of the m-version is suitable for
standard instruments by Gebrüder Martin. The socket can be used
to connect bipolar connection cables by Gebrüder Martin as well as
cables with US-2 flat connectors.
The socket of the socket module of the i-version is suitable for the
international instrument standard. The socket can be used to
connect bipolar connection cables having Erbe bipolar connectors
and US-2 flat-connectors.
The socket of the socket module of the e-version is suitable for the
international instrument standard. The socket can be used to
connect bipolar connection cables having Erbe bipolar connectors