maxium® smart C Electrosurgery Unit with maxium® smart Beam
124
Revision 1
17.2
Technical Data for maxium® smart C
Description
maxium® smart C
Mains connection
100–240 V ± 10%; 50/60 Hz
Mains power
max. 6.3 A
Mains fuse
6.3 A
Power consumption
without HF output
at maximum output power
45 VA
550 VA
Output power
Maximum cutting power
Maximum coagulation power
400 W
200 W
Protection class
I
Classification acc. to Medical Devices Act II b
Leakage currents LF and HF
according to IEC 60601-1 and IEC 60601-2-2
Type (Application Part)
CF; defibrillation-proof
Operating mode
Intermittent INT 10 s / 30 s corr. to 25% duty factor
Unit fuses
2 × T 6.3 AH, 250 V, Type G 5 × 20 mm
Signal level
HF display: 55 dB(A); adjustable 45–60 dB
Alarm: 65 dB(A)
Weight
9 kg
Radio interference suppression
The maxium® smart C meets the requirements of IEC 60601-1-2 also in
conjunction with the maxium® smart beam, but with the restrictions
pursuant to IEC 60601-2-2 Section 202.6.1 only during operation
readiness.
Dimensions
Width
Height
Depth
390 mm
160 mm
436 mm
Environmental conditions for transport
and storage
Ambient temperature
Relative humidity
Air pressure
-25°C to +70°C (-13°F to +158°F)
10–100%
500–1,060 hPa
Environmental conditions for operation
Ambient temperature
Relative humidity
Air pressure
+10°C to +40°C (50°F to +104°F)
30–75%
700–1,060 hPa
Compliant with 93/42/EEC
The technical data for the individual currents are to be found in the description of the individual currents in
section 11 “Currents, their Properties and Technical Data”, page 68.