Laser Surgical Unit diomax
®
70
Revision 6
See IEC 62353, Sections 5.4, for functional tests
[ ]
Control of calibration with an external power meter
(allowed tolerance ± 20%):
0.5 W / 2 W / 10 W / 15 W / 20 W
Additional test items:
Battery
[ ]
Check of internal battery (Ubatt
≥
2.8 V); replace battery if required
(CR2032 Lithium)
Error memory
[ ]
Select and retrieve internal memory “Event log” and record messages.
Take any required actions.
Emergency-Stop
[ ]
Check of EMERGENCY-STOP function
WARNING
If there are any safety-relevant problems after the safety check, the unitmust not be used
any longer.
Actions taken:
[ ] Technical Service contacted on: ..........................................................
[ ] Unit decommissioned on: ..................................................................
[ ] Results entered in medical device logbookon: ......................................
Inspector:
Date:
Signature: