Laser Surgical Unit diomax
®
Revision 6
51
5.12.6
Times
Up to four pulse parameters can be adjusted
depending on the pulse mode:
•
Pulse duration
(5 ms – 10 s)
Duration of one laser pulse
•
Pulse pause
(5 ms – 10 s)
Pause between two consecutive pulses in
pulse mode
•
Cycle
(20 ms – 40 s)
Total duration of one cycle consisting of
several pulses (at least 2) with pulse
pauses.
•
Cycle pause
(5 ms – 10 s)
Pause between two consecutive cycles
5.12.7
Saving the Modified Program
If the current program has been modified,
modified
is displayed next to the program
name.
Options for saving the modified program in the
menu
Programs
:
•
Save under
The modified program is saved under a new
name. The original will be stored
unchanged.
•
Overwrite
The original program will be replaced by its
modified version.
Save under
•
Select
Programs / Save under
—
The text editor will appear.