![KLS Martin diomax Instructions For Use Manual Download Page 68](http://html1.mh-extra.com/html/kls-martin/diomax/diomax_instructions-for-use-manual_1978994068.webp)
Laser Surgical Unit diomax
®
68
Revision 6
Description
diomax
®
data
Noise level
Noise at rest 42 dB(A), at full load 57 dB(A)
Dimensions (W × H × D)
381 mm × 195 mm × 390 mm
Weight
13.8 kg
Environmental conditions for
transport and storage
Ambient temperature
-25 °C to +50 °C
(-13 °F to +122 °F)
Relative humidity
(no condensing moisture!)
10 – 90%
Atmospheric pressure
700 – 1,060 hPa
Environmental conditions for
operation
Ambient temperature
+15 °C to +35 °C
(59 °F to 95 °F)
Relative humidity
(no condensing moisture)
30 – 75%
Atmospheric pressure
795 – 1,060 hPa
Safety check
annually
Compliant with 93/42/EEC