![KLS Martin diomax Instructions For Use Manual Download Page 67](http://html1.mh-extra.com/html/kls-martin/diomax/diomax_instructions-for-use-manual_1978994067.webp)
Laser Surgical Unit diomax
®
Revision 6
67
11
Tecnical Data
Description
diomax
®
data
Laser type
Diode laser
Laser wavelength
980 nm ± 2%
Laser output
0.1 – 20 W ± 20%
Pulse forms
Continuous mode:
Output:
0.1 – 20 W
Continuous rate:
Output:
1 – 20 W
Energy:
10 – 150 J/cm
Pulsed mode:
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Pulse rate:
Output:
1 – 20 W
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Energy:
10 – 150 J/cm
Cycle mode:
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Cycle time:
20 ms – 40 s
Cycle pause:
5 ms – 10 s
Single mode:
Output:
1 – 20 W
Pulse time:
5 ms – 10 s
Beam guidance
Laser fibers, focusing handpiece
Laser beam quality
Numerical laser aperture
≤
0.22
Laser fiber connection
SMA-X socket, mechanically coded SMA socket
Control and monitoring
2 microprocessors
Cooling
Semiconductor/air cooling
Mains connection
100 – 240 V ± 10%; 50 / 60 Hz
Mains power
max. 6.3 A
Mains fuse
2 x T6.3A H250V
Power consumption
650 W
Laser class
4
Protection class
I
Degree of protection
IP X1
Classification acc. to German
Medical Devices Act
II b
Pilot laser wavelength
635 nm
Pilot laser power
3 mW ± 20%, 2 – 100% adjustable, continuous and flashing
Pilot laser class
3R