4-14 Operation Method
RC04-PCR-LA
Table 4-4 Descriptions of Parameters of the Setting Display
• For items T5 or N, specify one
First
display
T1
Voltage change start time (phase)
Time or phase before starting the change.
Time or phase from the point that the waveform crosses the
zero level to the starting point of abnormal change, such as
voltage increase or decrease.
Select ms or deg as the unit.
For more information, see the Description on page 4-18.
T2
Slope time 1
Interval of time for the voltage to increase (pop) or
decrease (dip) from the initial value to the
fi
nal value spec-
i
fi
ed as abnormal voltage
T3
Voltage change duration
Interval of the time which elapses from the end of previous
voltage change to the beginning of the next voltage change
the voltage is maintained for Variation Voltage during the
interval.
Specifying zero(0) for this T3 disables this function.
Second
display
T4
Slope time 2
Interval of time for the voltage to increase (pop) or
decrease (dip) from the value speci
fi
ed as abnormal voltage
(Variation Voltage) to the initial value.
T5
Restoration duration
The time during the condition that the voltage recovered to
the initial value and has been maintained.
N
Restoration cycle
The number of cycles during the condition that the voltage
recovered to the initial value and has been maintained. The
number of cycles should be speci
fi
ed at current frequency.
V (T3)
Variation voltage
The
fi
nal voltage value increased (pop) or decreased (dip).
This value is speci
fi
ed at the interval of T3.
Third
display
RPT
Repetition count
The number of repetitions from T1 to T5 (N).