Turning On the Power
1
Inst
allation an
d Prep
aratio
n
Turning the POWER Switch On
When you turn the POWER switch on for the first time after purchase, the PBZ starts with its
factory default settings. Each time thereafter, the PBZ starts according to the setting that you
selected as outlined in “Function to perform when the POWER switch is turned on” on page 24.
1
Check that the power cord is connected properly.
2
Turn the POWER switch on.
Flip the POWER switch to the ( ) side to turn the PBZ on.
The following screen is displayed for approximately 2 seconds. At the same time, the
PBZ begins its self test.
•All keys illuminate, excluding the
CURSOR, SHIFT, and OUTPUT keys.
•The OUTPUT LED illuminates.
•The model name and firmware
version are displayed. When the self
test finishes, the PBZ starts. If no errors
were detected, “PASS” is displayed. If
an error was detected, the error
number is displayed.
Example: PBZ20-20 Firmware Ver.
1.00
If you notice strange sounds, unusual odors, fire, or smoke around or from inside the PBZ, flip
the POWER switch to the (
) side to turn the PBZ off, or remove the power cord plug from
the outlet.
When the PBZ Does Not Start Properly
The PBZ will not start properly if:
•
An error number is displayed after the self test.
•
An error number is displayed when you press the OUTPUT key.
•
“POWER OFF” is displayed.
If an error number is displayed after the self test
Follow the remedy that corresponds to the appropriate error number shown in the following
table. If following the remedy shown here does not solve the problem, contact your Kikusui
agent or distributor.
See
SELF TEST . . . PASS
SELF TEST . . . E101
When the test passes
When an error is detected
Error number
Message
Cause and remedy
E101
CALIBRATION DATA
An error was detected in the calibration values.
There is an error in the calibration values that are stored in
non-volatile memory. The PBZ must be calibrated properly.
If this error number (E101) appears even after you have
calibrated the PBZ, it may be malfunctioning.
E102
EEPROM DATA
An error was detected in the backup data.
There is an error in the backup data that is stored in non-
volatile memory. You can clear this error by starting the PBZ
with the factory default settings.
E901
INTERNAL ERROR
An error was detected in the internal system.
There is an execution error in the internal system. Restart
the PBZ.