7-16 Maintenance
PAS SERIES
■
Warming up
3.
Turn on the POWER switch while pressing the SET switch.
4.
Check that the voltage display shows "CAL."
If it is not, press the CURRENT switch.
5.
Warm up the equipment including the shunt resistor and
DVM for sufficient time.
■
OCP offset and full scale
6.
Press the OVP switch while pressing the SHIFT switch to
display "OCP" on the voltage display.
7.
Press the CURRENT switch to display "OF 0" on the cur-
rent display.
8.
Turn on the OUTPUT.
Calibration starts automatically (ON blinks) and the POWER
switch turns off. (This takes 60 s to 90 s.)
9.
Turn on the POWER switch while pressing the SET switch.
10.
Check that the voltage display shows "CAL."
If it is not, press the CURRENT switch.
11.
Press the OVP switch to display "OCP" on the current dis-
play.
12.
Press the CURRENT switch to display "FS 0" on the cur-
rent display.
13.
Turn on the OUTPUT.
You are done calibrating the OCP.