36 Remote Control
KDS6-0.2TR
V<n>Set
Sets the voltage value of Channel n, or inquires about the present setting. “n” should
be specified in a range from 1 to 3.
Program message
•
Syntax
Command message:
V<n>Set <Setting value>
V<n>S <Setting value>
Query message:
V<n>Set?
V<n>S?
•
Program data
(Example) To set the voltage value of Channel 1 to 3.1234 V:
V1Set 3.1234
Response message
In response to
V1S?
, the present setting is returned as fixed decimal point data
(0.0000 to 6.5000).
(Example)
3.1234
Setting of voltage value for Channel n
Minimum value
0.0000
Maximum value
6.5000
Resolution
0.0001
Data format
Fixed decimal point
Initial value
0.0000
Unit
V