KDS6-0.2TR
Remote Control 39
I3Out?
Inquires about the output current value of Channel 3.
Program message
•
Syntax
Query message:
I3Out?
I3O?
Response message
•
Response data
In response to
I30?
, the present output current value is returned as fixed deci-
mal point data.
If OCP has occurred, CH<n>OCP (n = 1 to 3) is returned, “n” indicating the
channel in which OCP took place. To clear OCP, the
OUTP
0
command is sent.
(Example 1)
30.0000
(Example 2)
CH3 OCP
IOUT?
Inquires about the output current values of Channels 1 to 3 collectively.
Program message
•
Syntax
Query message:
IOUT?
Response message
•
Response data
In response to
IOUT?
, the present settings of the channels are delimited with
“,” (ASCII:2Ch) and returned as fixed decimal point data. The data is returned
in the order CH1, CH2, CH3.
If OCP has occurred, CH<n>OCP (n = 1 to 3) is returned, “n” indicating the
channel in which OCP took place. To clear OCP, the
OUTP
0
command is sent.
(Example 1)
200.000,10.3245,-0.0005
(Example 2)
CH1 OCP,CH1 OCP,CH1 OCP
Output current reading resolutions for Channel 3
Resolution
Normal mode
0.0001
High-speed mode
0.001
Unit
mA
Output current reading resolutions for Channels 1 to 3
Resolution
Normal mode
0.001 (CH1 200-mA range)
0.0001 (CH1 10-mA range/CH2, 3)
High-speed mode
0.01 (CH1 200-mA range)
0.001 (CH1 10-mA range/CH2, 3)
Unit
mA
Summary of Contents for KDS6-0.2TR
Page 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Page 4: ...ii KDS6 0 2TR ...
Page 8: ...4 Safety Precautions KDS6 0 2TR ...
Page 12: ...8 Introduction KDS6 0 2TR ...
Page 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Page 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Page 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Page 58: ...54 Index KDS6 0 2TR ...
Page 59: ......