KDS6-0.2TR
Remote Control 33
*IDN?
Inquires about the model name of the precision DC source and the version of the
firmware.
Program message
•
Syntax
Query message:
*IDN?
Response message
In response to
*IDN?
, the model name, version no., etc., are returned as
shown in the example below:
ERR?
Inquires about the contents of an error register.
If an error occurs due to the content of a command message, this is reflected on each
bit of the error register. For a grammatical error, “1” is set at bit 0. If the program
header is correct, but there is an error in the program data portion, “1” is set at bit 1.
These contents are held until they have been read out of the error register. When
these contents have been read out, all bits are cleared to “0.”
Program message
•
Syntax
Query message:
ERR?
Response message
In response to
ERR?
, the present error code is returned.
(Example) When there is an error in the program data portion
1
Contents of error register
Bit
7
6
5
4
3
2
1
0
Error factor
–
–
–
–
–
–
Data
Error
Syntax
Error
Bit weight
128
64
32
16
8
4
2
1
KDS6-0.2TR,0, 1.00
Model name
Version No.
Company name
KIKUSUI ELECTRONICS CORP.,
Not in use
Summary of Contents for KDS6-0.2TR
Page 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Page 4: ...ii KDS6 0 2TR ...
Page 8: ...4 Safety Precautions KDS6 0 2TR ...
Page 12: ...8 Introduction KDS6 0 2TR ...
Page 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Page 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Page 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Page 58: ...54 Index KDS6 0 2TR ...
Page 59: ......