5
Service Guide
76
Keysight U8001A/U8002A User’s and Service Guide
4 High voltage calibration point.
– Display will show “C” blinking.
– The M3 annunciator will be turned on.
– Use the knob to enter the reading obtained from the DVM.
– Press the “Memory” button to save the changes voltage calibration is done.
You will be able to perform this if the power supply is in CV mode.
5 Voltage calibration
– Display will show “donE” for a few seconds and the power supply will enter
the current calibration mode.
NOTE
If you do not wish to continue with the current calibration, you may now turn off
the power supply to exit the calibration mode.