33
Keysight M9217A Startup Guide
-1.9 V
2 V
+/-0.0043 V
-121.6 V
128 V
+/-0.3776 V
0 V
+/-0.0024 V
0 V
+/-0.256 V
+1.9 V
+/-0.0043 V
+121.6 V
+/-0.3776 V
-3.8 V
4 V
+/-0.0158 V
-237.5 V
256 V
+/-0.5447 V
0 V
+/-0.012 V
0 V
+/-0.3072 V
+3.8 V
+/-0.0158 V
+237.5 V
+/-0.5447 V
-7.6 V
8 V
+/-0.0236 V
0 V
+/-0.016 V
+7.6 V
+/-0.0236 V
Table 4
Suggested calibrator output voltages
Input value
Channel range
Error from input
Input value
Channel range
Error from input
NOTE
The table uses the calibrator zero output for ease of connection. A more accurate
verification of the zero can be obtained by connecting a 50
Ω
termination to the input
connectors.
WARNING
Shock Hazard. This procedure requires the application of up to 240 Vdc. Ensure all test
conections are properly insulated and shielded.