Keysight M9217A Startup Guide
32
Figure 10
Data acquisition and measurement result
6
Compare the measured values to the accuracy as stipulated in the specifications.
7
for all voltage range and channels. Refer to the following table
for the list of suggested calibrator output voltages.
8
If any values are out of range, refer to “
for
details on sending the module to Keysight Technologies for service.
Table 4
Suggested calibrator output voltages
Input value
Channel range
Error from input
Input value
Channel range
Error from input
-235 mV
250 mV
+/-0.000985 V
-15.2 V
16 V
+/-0.0344 V
0 V
+/-0.00075 V
0 V
+/-0.0192 V
+235 mV
+/-0.000985 V
+15.2 V
+/-0.0344 V
-460 mV
500 mV
+/-0.00146 V
-30.4 V
32 V
+/-0.0688 V
0 V
+/-0.0010 V
0 V
+/-0.0384 V
+460 mV
+/-0.00146 V
+30.4 V
+/-0.0688 V
-950 mV
1 V
+/-0.00215 V
-60.8 V
64 V
+/-0.2528 V
0 V
+/-0.0010 V
0 V
+/-0.192 V
+950 mV
+/-0.00215 V
+60.8 V
+/-0.2528 V
Summary of Contents for M9217A PXIe 2-CH
Page 7: ...7 Keysight M9217A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 9: ...9 Keysight M9217A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 11: ...11 Keysight M9217A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 13: ...13 Keysight M9217A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 35: ...35 Keysight M9217A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...