62
InfiniiMax III+ Series Probes User’s Guide
2
Using InfiniiMax III+ Probe Heads
Figure 47
Probe Head Connection to DUT (
Non-InfiniiMode
)
Ensuring maximum performance
The specifications and performance plots of the N2836A probe head were
measured with a nominal spacing of 40 mil (1 mm). In order to achieve the proper
response as shown in the performance plots, keep the mini-axial lead
resistors
roughly parallel as shown in
, and use the tip wires on the mini-axial
leads to get the desired span.
Figure 48
Proper Position of Resistors
If you need to position the mini-axial lead resistors different than shown in this
figure (that is, resistor bodies close together or spread way apart), use N2807A
and N2808A PrecisionProbe products to perform an AC calibration of the probe.
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...