142
InfiniiMax III+ Series Probes User’s Guide
8
Performance Verification for N7000-Series Probes
b
On the VNA, go to menu “Calibration/Fixturing Selections/2 Port
De-embedding” and select Port 1.
c
Set S2P file selection to the file saved in step a.
d
Check the “Enable De-embedding”.
e
Under “Calibration” menu, select “Fixturing ON/off” to turn on
de-embedding.
Figure 102
Close-Up of Tip Wires Positioned Under Pincher
10
Trigger the VNA to perform a single sweep. Press “Trigger” under Channel
Setup, and then the green soft-key for “Single”. Display should look like
on page 143. If it looks noticeably different, the probe tip wires may
not be making contact under the pincher.
11
Under “Trace/Math/Memory” select “Data->Memory”. This will save the
de-embedded input voltage trace into the memory.