InfiniiMax III+ Series Probes User’s Guide
6
Performance Plots
Figure 86
Fidelity of Step Response (Probe System Versus Input Signal)
This example was for a 13 GHz system (N2832A probe) measuring a 6.6 GHz
signal. This ratio can be used to determine the measurement bandwidth needed
for other bandwidth signals (for a 3% rise time measurement error):
This equation can be used to determine the probe-scope system BW needed to
measure a certain BW signal.
BW
probe-scope system
13 GHz BW
signal
×
6.6 GHz
-------------------------------------------------
≅
BW
probe-scope system
1.97 BW
signal
×
≅