Using InfiniiMax III+ Probe Heads
2
InfiniiMax III+ Series Probes User’s Guide
53
Tip-span entry for probe calibration
When calibrating the probe, you will be prompted to enter the span setting on the
oscilloscope’s Probe Calibration dialog box. To determine the tip span, do the
following steps:
1
Adjust the browser’s tip span for your measurement.
2
Locate the tip-span gauge on the browser’s protective cap as shown in
3
Determine which of the three possible tip-span settings most closely matches
the browser’s tip span.
Figure 36
Tip Span Gauge on Browser Protective Cap
Mounting the Browser
There are two holes on the back side of the browser. Use these holes to mount the
browser to a customer designed holder.
below shows the dimensions of
these mounting holes.
Figure 37
Dimensions of Mounting Holes