Performance Verification for N7000-Series Probes
8
InfiniiMax III+ Series Probes User’s Guide
143
Figure 103
De-Embedded V
in
Trace
12
on page 144, move the Port 1 cable the N5477A.
Connect the 8775B2 broadband load as shown in the figure.
13
Use the following steps to Install a file to de-embed the adapter (5062-1247)
and N5477A from port 1 of the VNA.
a
Create the Touchstone file by cutting and pasting the text in
on page 129. Name the file
Adapter_5062_1247___Adapter_N5477A.s2p
. This is the same file that is used in
for the performance verification of N2830A-series probes.
b
Go to menu “Calibration/Fixturing Selections/2 Port De-embedding” and
select Port 1.
c
Set S2P file selection to the file saved in step a.
d
Make sure the “Enable De-embedding” box is still checked.
e
Under “Calibration” menu, make sure “Fixturing ON/off” is still checked so
file is being used for de-embedding.
NOTE
You can also copy the data from the Adobe AIR version of Keysight’s Probe Resource Center
(PRC). Copying this data from the PRC is the simplest most reliable method to get the data. To
download the PRC, visit http://www.Keysight.com/find/PRC.
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...