U3047AM08
Controlling the Test Set
Using Interface Control Mode
An Instrument Preset will reset all of the fields to their default settings.
Figure 30
Interface Control
Enable Interface Control: (1)
Enables and disables ALL Interface Control communication. When
cleared (default setting) Interface Control is disabled and NO data is sent.
To send data, the individual interfaces must also be enabled.
Test Set I/O Control (addr.data): (2)
Provides control of the test set I/O Interface on the rear panel of the network
analyzer. Used to control your test set.
Only positive integers are allowed to select switch positions or states of DUT
control interface lines. Refer to
“Address and Data Values” on page 4
Address and data integers must be separated by a period. Each set of entries
should be separated by a new line, or carriage return. The front panel Enter
key can be used to insert a new line into the field. The quantity of test set I/O
entries that can be entered is limited by the available memory of the network
analyzer. Refer
to the
.
If an error is encountered when sending Interface Control data, an error
message is displayed on the network analyzer screen and the Channel
Trigger State is set to Hold. You must fix the condition that caused the error,
then change the Channel Trigger State to its original setting.
(2)
(3)
(5)
(7)
(8)
(8)
(6)
(1)
(4)
40
Keysight U3047AM08 User's and Service Guide