U3047AM08
Control Lines
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
Setting the
N
etwork
A
nalyzer to Standalone Mode
1. Select Utility > System > Configure > Multiport Capability -OR-
(Instrument > Setup > External Hardware > Multiport > Multiport
Configuration...)
. In the dialog box, select
Restart as a standalone PNA > OK
.
The following control feature will function only while the network analyzer
is in its Standalone Mode.
Table 1
3
As described in
“Setting the Test Port Paths with Address and Data” on
, the <address>.<data> values are determined in a similar man
ne
r,
with the following exceptions:
—
will be used
— The <address> value is always
112
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8
Keysight U3047AM08 User's and Service Guide
53