Figure 5
4
Source IN to CPLR THRU & RCVR OUT to CPLR ARM Path Response
Table 2
1
U3047AM08
Service Information
Connect the RF Flex cables to the Source IN and CPLR THRU Port
Connect the RF Flex cables to the Source IN and CPLR THRU port indicated in
, the expected results should be similar to
.
Source Signal Path Test Instructions
Path #
Test Port
Group
RF Path Description
Control Mode
<Address>.<Data>
Path
Components
1
Port 5 & 9
Source IN to CPLR THRU
0.0
S10
2
Port 6 & 10
Source IN to CPLR THRU
16.0
S20
3
Port 7 & 11
Source IN to CPLR THRU
32.0
S30
4
Port 8 & 12
Source IN to CPLR THRU
64.0
S40