Using P-Series Power Sensor
3
Keysight N8262A User’s Guide
117
Figure 3-6
Trace display with trace control menu
The fields on the lower left of the screen are the X and Y trace setup fields. Using
the
or
softkeys you can highlight the item and change its value. (See
The table on the lower right of the screen shows the 7automatic time
measurements performed on the first complete captured pulse after the trigger.
The 7 measurements are rise time, fall time, time to positive occurrence, time to
negative occurrence, pulse width, pulse period (pulse repetitive interval) and
pulse repetitive frequency.
The current settings of the X and Y scale are displayed on the reporting line above
both tables.
Gate Control Click
to display the
Gate Ctrl
menu. Setting the
gate features are described in greater details in
“Gate control in a single enlarged
Trace Setting
Pulse Period and Pulse Width
Pulse Repetitive Frequency
Rising and
Falling Edge
Gate 2 Active
NOTE
If you want to view the trace in linear mode, the
Trace Setup
is the only location
where you can change the Y-scale units from
dBm
to
Watts
; otherwise the
default unit is in logarithmic.
Gate Control