Using P-Series Power Sensor
3
Keysight N8262A User’s Guide
111
Step 1. Connecting the power sensor
When you initially connect a P-Series Wideband Power Sensor to the power
meter, the sensor’s calibration data is automatically read by the power meter.
After the sensor’s data is read, the power meter automatically performs a zero and
calibration, see
“Zeroing and Calibrating the P-Series Sensor”
further information.
Connect the sensor to the power source.
Step 2. Setting the channel frequency
1
Click
. Click
to select the channel you want to configure. The
Channel Setup
screen is displayed.
2
Use the
,
,
, and
softkeys to highlight the frequency field.
3
Click
to configure the setting. See
on page 34 for further information.
Figure 3-1
P-Series sensor default channel setup
When you have completed the procedure you can return to this menu later to add
more accuracy to the measurement.
Channel