Maintenance
4
Keysight N8262A User’s Guide
127
Figure 4-1
Self test complete
When the test is complete, the result is displayed. Click
to return to the
Service
menu.
If the self test failed, information about the failure is displayed on the screen.
Remote testing
To invoke the remote self test, the IEEE 488.1 compliant standard command,
*TST?
is used. This command runs a full self test and returns one of the following
codes:
– 0 -no tests failed
– 1 -one or more tests failed
Done
NOTE
“Unterminated Query” and “Hardware Missing” error messages will be logged
during self test.
– “Unterminated Query” will be logged if the system is busy processing and no
response to P-Series Soft Front Panel application.
– “Hardware Missing” will be logged as the self test will disconnect the sensor
during certain stage self test.