General Power Meter Functions
102
Keysight N8262A User’s Guide
The
Zeroing
pop-up is displayed when a zero occurs.
Manual Zeroing
To manually zero the power meter and sensor:
Click
and the channel
softkey. Click
,
or
to start the zeroing routine.
The
Zeroing
pop-up is displayed.
Calibration
Calibration sets the gain of each power meter channel and P-Series sensor
combination. This is achieved without the need to connect it to the 1.0 mW Power
Reference. The power meter can be set to automatically calibrate
on-the-fly
while
you are using it. Alternatively, you can perform the calibration manually. The
reference calibration factor is automatically set for all P-Series Power Sensors.
During calibration, the wait symbol is displayed.
Figure 2-62
Calibration wait pop-up
Offset and relative settings are ignored during calibration.
NOTE
When you are performing measurements, this may cause delays in obtaining
results. During this time, it is recommended to disable the automatic zeroing.
Zero
Zero A Zero B
Zero Both