Keysight M9195A/B PXIe Digital Stimulus/Response Startup Guide
Step 5: Verify Operation of the M9195A/B Module
Figure 6
Connecting to Multiple Modules
If the Status indicator on the module front panel turns from green to blue, go on
If the module does not appear in the ‘Connect to Instrument” dialog box as
shown in
, go to step 2 of the troubleshooting checklist:
appear correctly on Keysight Connection Expert?
Then follow with
.
Conduct a Self Test
Now the M9195A/B is ready to run Self Test from the Soft Front Panel.
1
To start the Self Test, click
Utilities > Self Test...
and the
Run Self Test...
button.
Follow all screen prompts.
If M9195B modules are licensed for multi-module (option
M9195B-MMS) use they may be connected to the same SFP. The
can be connected to the SFP one at a time, or in a group as shown
in
.
Before you initiate the Self Test,
(a) Disconnect all cables (for accuracy and to prevent damage)
(b) Save any data that you need (waveforms, cycle instructions,
etc.), as Self Test will erase the memory in the module.
Self Test will take from 2 to 5 minutes to complete.