High-Level Diagnostic Tools, Processes and References
If the status LED does not go from solid green to solid blue after the
Soft
Front Panel
comes up and connects with the M9195A/B, there may be an
LED problem. If the
Soft Front Panel
does not come up, see
If the LED goes solid blue and the
Soft Front Panel
comes up, check that
the drivers are loaded correctly by flashing the blue LED from the SFP.
See
If an incorrect LED condition repeats and continues, go to
2
Does the M9195A/B appear correctly on
Keysight Connection Expert
?
–
Bring up
Connection Expert
and click on the “PXIe/AXIe Chassis” tab then
check to see if the M9195A/B in question appears in the correct slot.
If the module is not visible, try shutting down the system, cycling power,
and re-starting.
If the M9195A/B still does not show up in
Connection Expert
then
power down the chassis and repeat the process with the M9195A
/B
installed in another slot.
If the M9195A/B does not show up in
Connection Expert
, but the
Soft
Front Panel
if necessary, re-install all required software in the order describe in the
M9195A/B Startup Guide
.
If the Module continues to not show up in
Connection Expert
, go to
“Returning a module for service”
3
Does the M9195A/B (SFP) connect?
If the
Connection Expert
as well as the
SFP
does not connect, go through
Step 2 above.
–
If the SFP fails to connect to the M9195A/B, then
on page 30 and, if that is OK, try repeating the process with
the M9195A/B installed in another slot.
Re-install all required software in the order describe in the
M9195A/B Startup
Guide
.
–
If the SFP connects, go to
Summary of Contents for M9195A PXIe