![Keysight Technologies M9195A PXIe Startup Manual Download Page 30](http://html1.mh-extra.com/html/keysight-technologies/m9195a-pxie/m9195a-pxie_startup-manual_1961853030.webp)
Figure 10
SFP Automatic Corrections tab
24
Keysight M9195A/B PXIe Digital Stimulus/Response Startup Guide
Step 6: Execute Automatic Corrections
6
Select the
Automatic Corrections
tab
in the second row, as in
7
Click the
Update
button in
; this will update the status of whether
Automatic Corrections are required.
a
If the
Automatic Corrections Needed
field contains
False,
Automatic
Corrections are not needed; you can skip to the end of this procedure.
b
If the
Automatic Corrections Needed
field contains
True
, proceed to the next
step to execute the automatic corrections.
8
Select an
Averaging Mode
.
– Average64
is a good default if your system does not have power line noise.
–
Use
Window50Hz
if you notice power line noise and your local power is 50
Hz.
–
Use
Window60Hz
if you notice power line noise and your local power is 60
Hz.
9
Now, to run Automatic Corrections,
Click the
Execute Automatic Corrections...
button in
Summary of Contents for M9195A PXIe
Page 45: ......