Step 5: Verify Operation of the M9195A/B Module
Step 5: Verify Operation of the M9195A/B Module
In this step you will verify correct operation of the Keysight M9195A/B PXIe
Digital Stimulus/Response module. Before running Self Test or Automatic
Corrections, ensure that all required software is installed, the chassis is powered
on, and all front panel cabling is removed.
Check the Status LED
Check the M9195A/B front panel Status LED.
-
When power is first applied, the LED starts in the
Off
state and then changes
to
Green
a few moments after PXI enumeration takes place. Depending on
how the chassis is loaded with modules, this may take up to a few minutes.
-
In general, when the Status LED is solid green, the module is working
correctly.
-
table on
for the complete list of
the LED colors and their meaning.
If the LED does not go on after power is applied, see step 1 in the
troubleshooting checklist:
What does the front panel Status LED look like?
Check for M9195A/B with Connection Expert
Verify that Connection Expert recognizes the M9195A/B module.
To do this, after the host system boots up, bring up Connection Expert at the
Start Menu (see
below), and then enter “
M9195”
in the Filter
Instruments box and as shown in
below.
Figure 1
Selecting Connection Expert
Summary of Contents for M9195A PXIe