Keysight N9927-90001 User’s Guide 589
Noise Figure Mode (Option 356 – CPU2 Only)
How to use Auto-Integration
these cases, since the jitter goal has not been reached, a jitter warning marker
will appear on the data trace (unless disabled by the user), indicating that the
jitter goal could not be achieved within the selected time constraint.
Due to the stability of the FieldFox receiver with its internal reference switch
and the design of the auto integration feature, it is possible to achieve a stable,
accurate, low jitter measurement results even under challenging measurement
conditions, such as when measuring low noise power output devices at high
frequencies with no pre-amplifier, by setting very long maximum integration
times. For example, stable measurements with 0.1 dB jitter and no discernible
drift have been demonstrated measuring a low gain, low noise amplifier at
50 GHz using maximum integration times of 5000 seconds per point, in spite of
ambient temperature drift of several degrees Centigrade over the course of the
measurement.
While auto integration can achieve very low jitter measurements, it is a good
practice to consider the contribution of the jitter uncertainty to the total
measurement uncertainty when selecting a jitter goal. If the measurement
uncertainty from other contributors is known to be 0.5 dB, for example, it
would make little sense to set a jitter goal of 0.1 dB. Setting a jitter goal of
0.3 dB in this case would only increase the RSS uncertainty to 0.58 dB, and
would potentially speed up the measurement by a factor of 9 relative to a jitter
goal of 0.1 dB. Auto integration makes it possible to select the jitter level with
confidence, and is therefore a valuable tool in obtaining the required level of
accuracy with the fastest possible measurement time.
Auto-Integration Procedure
No user cal was performed for this measurement, because no external preamp
was used.
Set the equipment up in an area where it will remain undisturbed for the
determined amount of measurement time. Refer to
1.
Enter the Measurement type:
Select
Measure
>
Noise Figure
(default)
2.
Enter the Noise Source ENR information into a table:
Select
Measure Setup
Then choose:
—
Noise Source
then
—
ENR Mode Table
(default)
If you have previously saved your noise source’s ENR information into a
table, skip this step and proceed to step
.
Summary of Contents for FieldFox A Series
Page 6: ...6...
Page 596: ...596 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting...
Page 622: ...Data Analysis Features All about Scale Settings 622 Keysight N9927 90001 User s Guide...
Page 668: ...668 Keysight N9927 90001 User s Guide System Settings Service Diagnostics...
Page 718: ...718 Keysight N9927 90001 User s Guide Specifications Data Sheet...
Page 721: ...721 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation...
Page 722: ...722 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 723: ...Keysight N9927 90001 User s Guide 723 Korean Battery Safety Translation...
Page 724: ...724 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 725: ...Keysight N9927 90001 User s Guide 725 Korean Battery Safety Translation...
Page 726: ...726 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 727: ...Keysight N9927 90001 User s Guide 727 Korean Battery Safety Translation...
Page 728: ...728 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 729: ...Keysight N9927 90001 User s Guide 729 Korean Battery Safety Translation...
Page 730: ...730 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 731: ...731 Keysight Handheld Analyzers N99xxA User s Guide E Russian Battery Safety Translation...
Page 732: ...732 Keysight N9927 90001 User s Guide Russian Battery Safety Translation...
Page 733: ...Keysight N9927 90001 User s Guide 733 Russian Battery Safety Translation N9927 90001...
Page 734: ...734 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 735: ...Keysight N9927 90001 User s Guide 735 Russian Battery Safety Translation N9927 90001...
Page 736: ...736 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 737: ...Keysight N9927 90001 User s Guide 737 Russian Battery Safety Translation N9927 90001...
Page 738: ...738 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 739: ...Keysight N9927 90001 User s Guide 739 Russian Battery Safety Translation N9927 90001...
Page 740: ...740 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 741: ...Keysight N9927 90001 User s Guide 741 Russian Battery Safety Translation N9927 90001...
Page 742: ...742 Keysight N9927 90001 User s Guide Russian Battery Safety Translation...
Page 744: ...744 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA...
Page 745: ...Keysight N9927 90001 User s Guide 745 Keysight Software End User Licensing Agreement EULA...
Page 746: ...746 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA...
Page 754: ...Keysight N9927 90001 User s Guide Index...
Page 755: ......