192 Keysight N9927-90001 User’s Guide
SA (Spectrum Analyzer) Mode
SA Mode Settings
If the Detection method and Averaging type are set to Auto when you enable a
Band/Interval Power marker, the Detection method will change to Average
(RMS) and Averaging type will change to Power average. Other Detection
methods or Averaging type settings will usually cause measurement
inaccuracy. Learn more about Detection method in
, and Averaging Type in
Summary:
— In Zero span, an Interval Power Marker yields the average power over the
specified time interval.
— In non-Zero span, a Band Power Marker yields the total power over the
specified frequency range.
How to measure Band/Interval Power with a marker
— Press
Marker
to create or select a Normal or Delta marker to use to
measure Band/Interval Power.
— Then press
More
— Then
Marker Function
— Then
Band/Int Power
— Then
Band (or Interval) Span
— Then choose from the following:
—
Auto
Measures power over 5% of the displayed span, or ½ graticules.
The Band Span is automatically updated to remain at 5% when you
change the frequency span.
—
Man
Enter a frequency or time value using the
arrows or the
rotary knob, then press
Enter
. Or enter a value using the numeric
keypad and select a frequency or time suffix. The Band Span remains
at the frequency or time value that you set as the span changes.
Frequency Counter at Marker
Available in SA Mode ONLY.
Use any existing marker to make highly-accurate
frequency counter readings. For highest accuracy, lock the FieldFox to a stable
external frequency reference. Learn how in
“Frequency Reference Source” on
.
When Frequency Counter is ON, the FieldFox uses background sweeps to
zoom on the signal, measure the signal peak with 1 Hz resolution, and display
the frequency of the signal peak in the marker annotation area. The marker
does not move to the signal peak.
When Freq Counter is ON, measurement sweeps are considerably slower.
Summary of Contents for FieldFox A Series
Page 6: ...6...
Page 596: ...596 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting...
Page 622: ...Data Analysis Features All about Scale Settings 622 Keysight N9927 90001 User s Guide...
Page 668: ...668 Keysight N9927 90001 User s Guide System Settings Service Diagnostics...
Page 718: ...718 Keysight N9927 90001 User s Guide Specifications Data Sheet...
Page 721: ...721 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation...
Page 722: ...722 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 723: ...Keysight N9927 90001 User s Guide 723 Korean Battery Safety Translation...
Page 724: ...724 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 725: ...Keysight N9927 90001 User s Guide 725 Korean Battery Safety Translation...
Page 726: ...726 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 727: ...Keysight N9927 90001 User s Guide 727 Korean Battery Safety Translation...
Page 728: ...728 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 729: ...Keysight N9927 90001 User s Guide 729 Korean Battery Safety Translation...
Page 730: ...730 Keysight N9927 90001 User s Guide Korean Battery Safety Translation...
Page 731: ...731 Keysight Handheld Analyzers N99xxA User s Guide E Russian Battery Safety Translation...
Page 732: ...732 Keysight N9927 90001 User s Guide Russian Battery Safety Translation...
Page 733: ...Keysight N9927 90001 User s Guide 733 Russian Battery Safety Translation N9927 90001...
Page 734: ...734 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 735: ...Keysight N9927 90001 User s Guide 735 Russian Battery Safety Translation N9927 90001...
Page 736: ...736 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 737: ...Keysight N9927 90001 User s Guide 737 Russian Battery Safety Translation N9927 90001...
Page 738: ...738 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 739: ...Keysight N9927 90001 User s Guide 739 Russian Battery Safety Translation N9927 90001...
Page 740: ...740 Keysight N9927 90001 User s Guide Russian Battery Safety Translation N9927 90001...
Page 741: ...Keysight N9927 90001 User s Guide 741 Russian Battery Safety Translation N9927 90001...
Page 742: ...742 Keysight N9927 90001 User s Guide Russian Battery Safety Translation...
Page 744: ...744 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA...
Page 745: ...Keysight N9927 90001 User s Guide 745 Keysight Software End User Licensing Agreement EULA...
Page 746: ...746 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA...
Page 754: ...Keysight N9927 90001 User s Guide Index...
Page 755: ......