background image

 

 5

 

 

• Ground the instrument

To avoid electric shock hazard, the instrument chassis and cabinet must be 
grounded by using the supplied power cable’s grounding prong.

• DO NOT operate in an explosive atmosphere

Do not operate the instrument in the presence of flammable gasses or fumes. 
Operation of any electrical instrument in such an environment clearly 
constitutes a safety hazard. 

• Keep away from live circuits

Operators must not remove instrument covers. Component replacement and 
internal adjustments must be made by qualified maintenance personnel. Do not 
replace components with the power cable connected. Under certain conditions, 
dangerous voltages may exist even with the power cable removed. To avoid 
injuries, always disconnect power and discharge circuits before touching them.

• DO NOT service or adjust alone

Do not attempt internal service or adjustment unless another person, capable of 
rendering first aid and resuscitation, is present.

• DO NOT substitute parts or modify instrument

To avoid the danger of introducing additional hazards, do not install substitute 
parts or perform unauthorized modifications to the instrument. Return the 
instrument to a Keysight Technologies Sales and Service Office for service and 
repair to ensure that safety features are maintained in operational condition.

• Dangerous Procedure Warnings

Warnings, such as the example below, precede potentially dangerous 
procedures throughout this manual. Instructions contained in the warnings must 
be followed.

Dangerous voltages, capable of causing death, are present in this instrument. 
Use extreme caution when handling, testing, or adjusting this instrument.

Certification

Keysight Technologies certifies that this product met its published specifications at 
the time of shipment from the factory. Keysight Technologies further certifies that 
its calibration measurements are traceable to the United States National Institute of 
Standards and Technology, to the extent allowed by the Institution’s calibration 
facility, or to the calibration facilities of other International Standards Organization 
members.

Summary of Contents for 16034G

Page 1: ...Keysight 16034G Test Fixture Operation Manual...

Page 2: ......

Page 3: ...3...

Page 4: ...in reprints do not necessitate a new printing date The manual part number changes when extensive technical changes are incorporated January 1999 Edition 1 part number 16034 90001 April 1999 Edition 2...

Page 5: ...DO NOT substitute parts or modify instrument To avoid the danger of introducing additional hazards do not install substitute parts or perform unauthorized modifications to the instrument Return the i...

Page 6: ...rrants that its software and firmware designated by Keysight Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Keysight Techno...

Page 7: ...upply Off Supply In position of push button switch Out position of push button switch Frame or chassis terminal A connection to the frame chassis of the equipment which normally includes all exposed m...

Page 8: ...8...

Page 9: ...ng the 16034G 16 Performing Short Correction 17 Performing Open Correction 18 DUT Measurement 19 DUT measurement over 3 MHz 20 3 Specifications and Supplemental Performance Characteristics Specificati...

Page 10: ...10 Contents...

Page 11: ...1 Overview...

Page 12: ...erview Product Overview Product Overview The 16034G is designed for chip type components whose sizes range from 0201 to 1306 This test fixture can take measurements of the chip type L C R Figure 1 1 P...

Page 13: ...shipment should be as listed in Table 1 1 If the contents are incomplete if there is mechanical damage or defect notify the nearest Keysight Technologies office If the shipping container is damaged or...

Page 14: ...nstrument s LCUR LPOT and HIGH side Contact Pin connected to a instrument s HCUR HPOT 2 Knob For lateral adjustment of HIGH side Contact Pin 3 Contact Pin Locking Screw For securing HIGH side Contact...

Page 15: ...2 Operation This chapter describes the proper methods for open and short correction and DUT measurement...

Page 16: ...e shows correction and measurement by the 16034G Connecting the 16034G 1 Set the cable length to 0 m in the instrument 2 Connect the 16034G directly to the UNKNOWN terminals as shown in Figure 2 1 Fig...

Page 17: ...llows 1 Push the HIGH side contact pin s knob to the left to make firm contact with the LOW side contact pin Figure 2 2 Tighten the contact pin locking screw to secure the HIGH side contact pin Figure...

Page 18: ...recommended that you place the DUT on the measuring stage and precisely position the HIGH side contact pin to actual DUT width 2 Tighten the contact pin locking screw to secure the HIGH side contact...

Page 19: ...of the measuring stage and secure the contact pin with the contact pin locking screw 2 Release the LOW side contact pin with the lever and set the DUT on the measuring stage 3 Ease back slowly on the...

Page 20: ...DUT measurement and perform measurement at 3 MHz to determine the value of the 100 SMD resistor 2 Set the measured resistance and inductance values to the instrument as load value 3 Perform load corre...

Page 21: ...3 Specifications and Supplemental Performance Characteristics This chapter provides specifications and supplemental performance characteristics of the 16034G test fixture...

Page 22: ...nalyzers with four terminals Applicable DUT Type Chip components Applicable DUT dimensions Maximum Voltage 40 V peak max AC DC Operating Environment temp 0 C to 55 C humidity 15 to 95 RH wet bulb temp...

Page 23: ...ORT correction Additional errors are calculated as follows Z Measurement Additional error Ze of the Z measurement is calculated by substituting the values in the table below into the following equatio...

Page 24: ...l error Rse of the Rs measurement is calculated by additional error Ze of Z measurement as follows If Dx 0 1 Rse Ze Dx If 0 1 Dx 0 5 Rse Ze Dx Dx is the measured value of D and is calculated as follow...

Page 25: ...ics Supplemental Performance Characteristics Contact Pressure The following data are supplemental performance characteristics for the spring that applies contact pressure Spring constant 37 gf mm 10 S...

Page 26: ...26 Chapter 3 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics...

Page 27: ...This information is subject to change without notice Keysight Technologies 1999 2001 2015 2016 Edition 6 November 2016 16034 90011 16034 90011 www keysight com...

Reviews: