24
Chapter 3
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
where Dx is the measured value of D. It is necessary for Ze to be below 10 %.
Rs (ESR) Measurement
Additional error Rse[%] of the Rs measurement is calculated by additional error
Ze [%] of |Z| measurement as follows.
If Dx
0.1:
Rse [%] = Ze / Dx
If 0.1
Dx
0.5:
Rse [%] = ( Ze
/ Dx) ×
Dx is the measured value of D and is calculated as follows.
Dx = 2 ×
× f × Csx × Rsx,
where
f: measurement signal frequency
Csx: measured value of Cs
Rsx: measured value of Rs.
Figure 3-1
Additional Error in |Z| measurement
D is not expressed as a percentage but as an absolute value.
1 Dx
2
+