ALJ/ALS/PLJ/PLS-BA-e-1135
11
KERN
PLS 3100-2F
PLS 4200-2F
PLS 4200-2A
Weighing range (max)
3100 g
4200 g
4200 g
Readability (d)
0.01 g
0.01 g
0.01 g
Reproducibility
0.02 g
0.02 g
0.01 g
Linearity
± 0.03 g
± 0.04 g
± 0.02 g
Stabilization time (typical)
3 sec.
3 sec.
2 sec.
Minimum unit weight at
piece counting
> 50 mg
> 50 mg
> 50 mg
Warm-up time
2 hours
2 hours
4 hours
Recommended adjustment
weight, not added (class)
3 kg (E2)
4 kg (E2)
4 kg (E2)
Reference quantities at
piece counting
10, 25, 50, 100 freely selectable
Weighing Units
ct, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw), pen
Electric Supply
230V/50 Hz (Euro)9V DC
230V/50 Hz
(Euro)24V AC
Rechargeable battery
Operation period 30 h
Loading time 10 h
no
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Underfloor weighing
Clevis type eyelet, standard
Housing (B x D x H) mm
210 x 340 x 160 210 x 340 x 100 210 x 340 x 100
Windshield mm
inside Ø 150, 60 high
outside Ø 160, 70 high
no
no
no
Weighing plate
(stainless steel)
Ø 16 cm
Ø 16 cm
Ø 16 cm
Weight kg (net)
4 kg
4 kg
4.5 kg
Interface
RS 232C