ALJ/ALS/PLJ/PLS-BA-e-1135
72
12 Error messages
ERR01
Weight value instable or zeroing not possible. Check the
environmental conditions.
ERR02
Adjustment error, e.g. environmental conditions instable
ERR03
Adjustment error e.g. incorrect adjustment weight}
ERR04
Piece weight too small/instable
ERR05
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
ERR06
Weighing value in density determining mode instable.
Check the environmental conditions.
ERR07
Incorrect data reading
(piece counting, density determination,…)
ERR08
Error at internal adjustment
“
UNLOAD
”:
Weighing range not achieved.
Check position of the weighing plate
“
CAL But
”
Adjust balance
Weighing range exceeded, placed load exceeds the capacity of
the balance. Unload the balance.
Weighing range not reached, e.g. weighing plate not in place