7.2 Models with internal weight (KERN ALJ/PLJ)
For non verified balances four adjustment possibilities are available in the menu.
Menu settings
In weighing mode press the
MENU
button and keep it pressed until
the acoustic signal gets mute. The first menu item „units“ is displayed.
Press
MENU
button repeatedly until „calib“ appears and confirm using
button.
Use the
MENU
key to choose between the following settings:
AUT-CAL
Automatic adjustment with internal weight.
Factory settings of verified balances.
I-CAL
Adjustment with internal weight after having pressed
CAL, locked in verifiable units.
E-CAL
Adjustment with external weight
locked for verifiable units
TEC-CAL
not documented
Take over selection using the
button.
The balance returns to menu.
To finish the menu press the
MENU
button and keep it pressed until
the acoustic signal gets mute. The balance returns automatically into
weighing mode.
AUT-CAL
Factory setting of
verified balances
With activated AUT-CAL function the internal adjustment is
automatically started when the balance
•
after the weighing balance was disconnected from
the mains
•
after pressing
ON/OFF
in stand-by mode
•
after a temperature change of 1.5 °C
with non loaded weighing plate / zero display
•
after a time interval of 20 min
with non loaded weighing plate / zero display
The automatic adjustment function is always enabled. You can
start adjustment at any time by pressing the
CAL
-key manually.